The NeuroFortis is an automatic, motorized, washable drive. Compatible with all Alpha Omega systems.

Key Features

  • Compatible with Automated Washing Machine*
  • Meets Robert Koch Institute Recommendations for Cleaning and Sterilization of Medical Devices   
  • Preassembled with built-in recording and stimulation capability
  • Operated by remote control or manually
  • Drive with micrometer resolution feedback for precision movement
  • High amplifier input range allows non-stimulating channels to record throughout stimulation
  • Signal digitization and broadband filtering
  • 10 channels capable of recording unfiltered raw, high-frequency spike and low-frequency LFP signals**
  • Drives 5 microelectrodes simultaneously
  • New ergonomic knob design for user-friendly and accurate manual driving
  • A single cable connects the drive to all electrodes
  • A Single cable crossing the sterile field for all channels recording, and stimulation
  • High-quality material, durable design and production control for maximum durability in aggressive OR environment
  • XY stage and 5-hole Bengun allows trajectory adjustments without frame manipulation
  • Direct implantation of DBS electrode without altering the position of the drive or removing the cannulas
  • Lightweight, low impact on frameless or frame based procedure
  • Fully compatible with all stereotactic frames or frameless procedure

* Please follow the reprocessing instructions

**Available for NeuroOmega only. NeuroSmart is capable of recording 5 channels high-frequency spike and 5 channels low-frequency LFP signals no raw recording.

 

Disclaimer: This page may include information about products that may not be available in your region or country. Please consult the approved indications for use. Content on specific Alpha Omega products is not intended for users in markets that do not have authorization for use.

Want to hear more about NeuroFortis?

Fill up your details and we will get back to you within 4 business days

Error
Order now

Range

40 mm

Resolution

100µm

Dynamic Input range

1mV

Input impedance

100MΩ/30pF

Frequency response (LFP Channels)

1Hz - 250Hz

Frequency response (Spike Channels)

400Hz-5.5KHz

Noise (1 Hz- 10KHz BW)

<8V P-P when inputs are shorted to GND

Input power

5V DC and 18V DC

Cleaning and Sterilization

Washable, Autoclavable

Case dimensions

110mm X 40mm X 50mm

Weight

190 Grams

The NeuroFortis can be used with various frame adaptors to fit a range of  Stereotactic frames,
The frame adaptor fits onto the Stereotactic frame and serves as the base on which to attach the NeuroFortis Drive Headstage.

  • Leksell frame
  • Leibinger ZD & RM Frames
  • CRW Frame
  • Nexframe
  • Starfix Frame
  • ROSA ONE Brain- robotic System
  • Renishaw – Neuromate Robotic System